    • AO3051 high-precision monocrystalline silicon differential pressure/pressure transmitter

    Standard specification for monocrystalline silicon differential pressure transmitter

     

    (Calibration range based on standard zero point, stainless steel 316L diaphragm, filled with silicone oil)

     

    Performance specifications

     

    Reference accuracy for range adjustment

     

    (including linearity, hysteresis, and repeatability from zero) ± 0.075%

     

    If TD>10 (TD=large range/adjustable range), it is: ± (0.0075 × TD)%

     

    The square root output accuracy is 1.5 times the linear reference accuracy mentioned above

     

    1. Environmental temperature impact

    Range code

    -20℃~65℃total impact

    A

    ±(0.45×TD+0.25)%×Span

    B

    ±(0.30×TD+0.20)%×Span

    C/D/E

    ±(0.20×TD+0.10)%×Span

    Range code

    -40℃~-20℃和65℃~85℃total impact

    A

    ±(0.45×TD+0.25)%×Span

    B

    ±(0.30×TD+0.20)%×Span

    C/D/E

    ±(0.20×TD+0.10)%×Span

    Over range impact: ± 0.075% x Span

    2. Static pressure influence

    Range code

    influence quantity

    A

    ±(0.15%URL+0.10%Span)/4MPa

    B

    ±(0.10%URL+0.075%Span)/16MPa

    C/D/E

    ±(0.05%URL+0.05%Span)/16MPa

    3. Overvoltage impact

    Range code

    influence quantity

    A

    ±0.2%×Span/4MPa

    B

    ±0.2%×Span/16MPa

    C/D/E

    ±0.1%×Span/16MPa

    4. Long term stability

    Range code

    influence quantity

    A

    ±0.5%×Span/1year

    B

    ±0.2%×Span/1year

    C/D/E

    ±0.1%×Span/1year

    4. Long term stability power supply impact: ± 0.001%/10V (12-42V DC), negligible.

     

    Functional specifications


     

    1. Range and Range

    range/purview

    kPa

    mbar

    A

    range

    0.1~1

    1~10

    purview

    -1~1

    -10~10

    B

    range

    0.2~6

    2~60

    purview

    -6~6

    -60~60

    C

    range

    0.4~40

    4~400

    purview

    -40~40

    -400~400

    D

    range

    2.5~250

    25~2500

    purview

    -250~250

    -2500~2500

    E

    range

    20~2000

    0.2~20 bar

    purview

    -500~2000

    -5~20bar

    2. Range upper and lower limits


     

    Within the upper and lower limits of the measuring range, it can be adjusted arbitrarily.


     

    Suggest selecting range codes with the lowest possible range ratio to optimize performance characteristics.


     

    3. Zero point setting


     

    The zero point and range can be adjusted to any value within the measurement range in the table, as long as the calibration range is ≥ small range


     

    4. Installation location impact


     

    Changes in the installation position parallel to the membrane surface will not cause zero drift effects. If the installation position changes by more than 90 degrees from the membrane surface, there will be zero position effects within the range of<0.4kPa, which can be corrected by adjusting the zero position. No range influence.


     

    5. Output


     

    2-wire system, 4-20mADC, optional HART output digital communication, optional linear or square root output.


     

    Output signal limit: Imin=3.9mA,Imax=20.5mA


     

    6. Alarm current


     

    Low alarm mode (small): 3.7 mA


     

    High alarm mode (large): 21 mA


     

    Non reporting mode (hold): Maintain the effective current value before the fault


     

    Alarm current standard setting: high alarm mode


     

    7. Response time


     

    The damping constant of the amplifier component is 0.1s; The time constant of the sensor is 0.1 to 1.6 seconds, depending on the range and range ratio. The additional adjustable time constant is 0.1 to 60 seconds.


     

    The impact on nonlinear outputs (such as square root function) depends on the function and can be calculated accordingly.


     

    Preheating time:< 15s


     

    Environmental temperature: -40~85 ℃; -20~65 ℃ with LCD display and fluororubber sealing ring


     

    Storage temperature/transportation temperature: -50~85 ℃; When equipped with LCD display:- 40~85℃


     

    Working pressure: The rated working pressure is divided into three levels: 16MPa, 25MPa, and 40MPa


     

    Static pressure limit: From a pressure of 3.5kPa to the rated pressure, the protective pressure can be greater than 1.5 times the rated pressure and applied to both sides of the transmitter.


     

    Unidirectional overload limit: Unidirectional overload can reach the rated pressure


     

    Electromagnetic Compatibility (EMC)


     

    Refer to the "Electromagnetic Compatibility Appendix" on the next page


     

    Installation


     

    1. Power supply and load conditions


     

    The power supply voltage is 24V, R ≤ (Us-12V)/Imax k Ω; where Imax=23 mA


     

    Large power supply voltage: 42VDC


     

    Small power supply voltage: 12VDC, 15VDC (backlit LCD display)


     

    Digital communication load range: 250~600Ω


     

    2. Electrical connection


     

    M20X1.5 cable sealing buckle, the terminal block is suitable for wires of 0.5-2.5mm2.


     

    3. Process Connection


     

    The two ends of the process connection flange have NPT 1/4 and UNF 7/16 internal threads.


     

    Physical specifications


     

    1. Material


     

    Measurement membrane box: stainless steel 316L


     

    Diaphragm: Stainless steel 316L, Hastelloy C


     

    Process flange: Stainless steel 304


     

    Nut and bolt: stainless steel (A4)


     

    Filling fluid: Silicone oil


     

    Sealing rings: Nitrile rubber (NBR), Fluororubber (FKM)


     

    Polytetrafluoroethylene (PTFE)


     

    Transmitter housing: made of aluminum alloy material, coated with epoxy resin on the outside


     

    Shell sealing ring: Nitrile rubber (NBR)


     

    Nameplate: Stainless Steel 304


     

    Weight: 3.3kg (none: LCD display, mounting bracket, process connection)


     

    Shell protection level: IP67


     

    2. Electromagnetic Compatibility Appendix

    Serial Number

    Test project

    Basic standards

    Test conditions

    Performance Level

    1

    Radiation interference (shell)

    GB/T 9254-2008Table 5

    30MHz~1000MHz

    qualified

    2

    Conducted interference


     

    (DC power port)

    GB/T 9254-2008Table 1

    0.15MHz~30MHz

    qualified

    3

    Electrostatic discharge (ESD) immunity

    GB/T 17626.2-2006

    4kV(contact)

    8kV(air)

    B

    4

    Radio frequency electromagnetic field immunity

    GB/T 17626.3-2006

    10V/m (80MHz~1GHz)

    A

    5

    Power frequency magnetic field immunity

    GB/T 17626.8-2006

    30A/m

    A

    6

    Electrical fast transient burst immunity

    GB/T 17626.4-2008

    2kV(5/50ns,5kHz)

    B

    7

    Surge immunity

    GB/T 17626.5-2008

    1kV(Between lines)

    2kV(Between lines and ground)(1.2us/50us)

    B

    8

    Immunity to Conducted Interference Induced by Radio Frequency Field

    GB/T 17626.6-2008

    3V (150KHz~80MHz)

    A

     

  • AO3051DP/GP isolated intelligent remote differential pressure/pressure transmitter
    Instrument ModelAO3051DPisolated intelligent remote differential pressure/pressure transmitter
    Measuring RangeB-6 kPa~6 kPa
    C-40 kPa~40 kPa
    D-100 kPa~250 kPa
    F-0.1 MPa~3 MPa
    G-0.1 MPa~10 MPa
    Structural Material                 Flanges and JointsIsolation Diaphragm
    22316(Stailess Steel)316L(Stailess Steel)
    23Hastelloy C
    25Tantalum (above 6kPa)
    26Gold Plating
    27PTFE Spraying
    28Titanium
    29Nickel
    00Other Materials
    Remote Transmission DeviceS1one remote transmission device
    S2two remote transmission device
    Basic ErrorD±0.1% (above 6kPa, ranges B~G optional)
    E±0.2% (above 6kPa, ranges B~G optional)
    F±0.5% (above 6kPa, ranges B~G optional)
    Pressure Chamber Flange; Fastener MaterialEGalvanized Carbon Steel
    FStainless Steel 304
    GStainless Steel 316
    Optional PartsB1Pipe-Mounted Elbow Bracket
    B2Panel-Mounted Angled Bracket
    B3Pipe-Mounted Flat Bracket
    SUStainless Steel 316 Housing
    XNo Oil Allowed
    YNo Water or Oil Allowed
    ZDegreased
    iIntrinsically Safe Type Ex ia IIC T6
    dFlameproof Type Ex d IIC T6
    Selection ExamplesAO3051DP - C - 22 - S2 - D - E - B1

Message

If you have any needs, please feel free to contact us at any time.

Submit Now

Related Products

message

HOW CAN WE HELP ?

Get In Touch With Us!

%{tishi_zhanwei}%

Contact Us

Address:

No.36, Longhui Road, Wujin National Hi-Tech Industrial Development Zone, Jiangsu Province, China

Address:

Building 5 and 11, Fuzhou IoT Industrial Park, 

No. 12 Xingye West Road, Mawei District, Fuzhou City

Tel:

+86-13055728858

Email:

aotai03@fjautec.com      Mr. Zhou

aotai01@fjautec.com       Mr. Liao

aotai02@fjautec.com      Mr. Zhang

aotai04@fjautec.com      Mr. Chen

Email:

autec@163.com


Copyright © 2025 Fujian Aotai Automation Equipment Co., Ltd.,  All Rights Reserved.